The single-port concept: combining optical glucose measurement with insulin infusion.

Publikation aus Health

Hajnsek M , Nacht B, Sax S, List EJ, Klimant I, Sinner F.

Acta Diabetologica, 2014 Oct;51(5):883-6, 10/2014

Abstract:

The aim of this study was to develop a single-port system that combines glucose monitoring and insulin infusion. In a preclinical proof-of-concepttrial, the performance of the glucose measurement at the site of insulin infusion was assessed. Glucose levels were clamped from 40 to 250 mg/dL by intravenous glucose infusion and subcutaneous insulin infusion via the glucose sensor. Sensor-glucose values correlated well with reference blood-glucose values, despite infusion at the site of glucose measurement. The average median ARE value was 21.6 ± 5.7 % for sensors used forinsulin infusion, 18.1 ± 5.8 % for sensors used for NaCl infusion and 19.2 ± 7.9 % for sensors without infusion. These preclinical in vivo results demonstrate that single-port glucose monitoring is feasible at the site of insulin infusion.